Back to search
FDAFDA · Food and drugs·Recall number Z-2385-2012

Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine.

Published September 26, 2012 · Recall initiated: June 11, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Product sterility may be compromised due to a mechanical failure of the protective packaging tube. Tube end plugs may dislodge creating the potential for the screw to come out of the protective tube and then the sharp end of the screw pierce the outer sterile barrier pouch.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Tedan Surgical Innovations Llc
Distributor
Worldwide distribution: USA (nationwide) and countries of : Australia and Taiwan.
Lot
Lot numbers: 019715, 019716, 019942, 019943, 019977, 019978, 020102, 020103, 020104, 020105, 020106, 020107, 020108, 020609, 020782, 021539, 021663, 022097, 022098, 022099, 022122, 022123, 022124, 022371, 022372, 022790, 022791, 022792, 022794, 022795, 022796, 022827, 022828, and 023266.
Risk classification
Class II
Area (FDA)
device
Country
United States