FDAFDA · Food and drugs·Recall number Z-2445-2024

LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

Published July 31, 2024 · Recall initiated: June 27, 2024

Summary

FDA records medical device Class II recall Z-2445-2024, published July 31, 2024 and initiated June 27, 2024. Recalling firm: LINK BIO CORP; Status: Ongoing; Reported quantity: 8 units; Lot: Item Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
LINK BIO CORP
Distributor
US Nationwide distribution in the states of FL, GA, KS, OH, WI.
Lot
Item Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00 recalled?

Yes. FDA published recall Z-2445-2024 on July 31, 2024. Reported hazard: The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.