LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00
Published July 31, 2024 · Recall initiated: June 27, 2024
Summary
FDA records medical device Class II recall Z-2445-2024, published July 31, 2024 and initiated June 27, 2024. Recalling firm: LINK BIO CORP; Status: Ongoing; Reported quantity: 8 units; Lot: Item Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- LINK BIO CORP
- Distributor
- US Nationwide distribution in the states of FL, GA, KS, OH, WI.
- Lot
- Item Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPotassium Chloride Extended-Release Capsules, USP, (750 mg) 10 mEq K, a).100-count bottle ( NDC 68001-396-00),August 7, 2024
- FDAUmary Acido Hialuronico, Suplemento Alimenticio, 850 mg caplets, packaged in 30-count bottles.August 7, 2024
- FDAGlenmark, Azelaic Acid Gel, 15 %, 50 grams, Rx only, Manufactured by: Glenmark Pharmaceuticals Limited,August 7, 2024
- FDASensis Vibe systems with software version VD15B-Intended to be used as a diagnostic and administrative toolJuly 31, 2024
- FDASensis Vibe systems with software version VD15B in combination with AXIOM Sensis Post ProcessingJuly 31, 2024
- FDASensis Vibe systems with software version VD15B in combination with VM VIRTUAL SERVER Model: 10765502-IntendedJuly 31, 2024
Questions about this recall
Was LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00 recalled?
Yes. FDA published recall Z-2445-2024 on July 31, 2024. Reported hazard: The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail