Zimmer Gender Solution Patello-Femoral Joint (PFJ) System Patello-Femoral Trochlea Component, Precoat, Size 4, Left. Knee prosthesis.
Published September 3, 2014 · Recall initiated: July 30, 2014
Summary
FDA records medical device Class II recall Z-2545-2014, published September 3, 2014 and initiated July 30, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 15 units; Lot: 00-5926-014-01; Lot 62455332. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to final manufacturing under Part 00-5926-014-01; lot 62455332. A forged lap is a defect that forms when metal folds over itself during forging.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - US (nationwide) in the states of HI, CA, GA, MI, and IL and the countries of TAIWAN, BELGIUM, AUSTRALIA, FRANCE, JAPAN, and UK.
- Lot
- 00-5926-014-01; Lot 62455332
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E,September 3, 2014
- FDAKnee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESSSeptember 3, 2014
Questions about this recall
Was Zimmer Gender Solution Patello-Femoral Joint (PFJ) System Patello-Femoral Trochlea Component, Precoat, Size 4, Left. Knee prosthesis. recalled?
Yes. FDA published recall Z-2545-2014 on September 3, 2014. Reported hazard: Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to final manufacturing under Part 00-5926-014-01; lot 62455332. A forged lap is a defect that forms when metal folds over itself during forging.
What should I do if I have this product?
Voluntary: Firm initiated · Letter