CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.
Published September 3, 2014 · Recall initiated: August 5, 2014
Summary
FDA records medical device Class II recall Z-2554-2014, published September 3, 2014 and initiated August 5, 2014. Recalling firm: Curbell Medical, Inc.; Status: Terminated; Reported quantity: Domestic: 112 units; Foreign: 211 units; Lot: Lot 0414. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Curbell Medical, Inc.
- Distributor
- Nationwide, Canada, Ireland, Italy
- Lot
- Lot 0414
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASiemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety ofSeptember 3, 2014
- FDAThe Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, andSeptember 3, 2014
- FDAKnee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM IMPLANT 22MMDX145MM STEM IMPLANTSeptember 3, 2014
- FDACordis EMPIRA NC RX PTCA Dilatation CatheterSeptember 3, 2014
- FDAMoore & Thompson Hip Stems: FNSTRD HIP 38HDX127ST FNSTRD HIP 40HDX127ST FNSTRD HIP 41HDX127ST FNSTRD HIPSeptember 3, 2014
- FDACordis EMPIRA RX PTCA Dilatation CatheterSeptember 3, 2014
Questions about this recall
Was CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm. recalled?
Yes. FDA published recall Z-2554-2014 on September 3, 2014. Reported hazard: The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail