The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of used irrigant.
Published September 3, 2014 · Recall initiated: February 10, 2014
Summary
FDA records medical device Class II recall Z-2555-2014, published September 3, 2014 and initiated February 10, 2014. Recalling firm: Thermedx LLC; Status: Terminated; Reported quantity: 21; Lot: Model P4000 Serial numbers affected: 20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Thermedx LLC
- Distributor
- Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.
- Lot
- Model P4000 Serial numbers affected: 20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUALSeptember 3, 2014
- FDACordis EMPIRA NC RX PTCA Dilatation CatheterSeptember 3, 2014
- FDASiemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety ofSeptember 3, 2014
- FDACordis EMPIRA RX PTCA Dilatation CatheterSeptember 3, 2014
- FDAKnee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM IMPLANT 22MMDX145MM STEM IMPLANTSeptember 3, 2014
- FDACollectEVAC, Sterile device with tubing and adaptor for use with ACMI and STORZ/Wolf resectoscopes. PartSeptember 3, 2014
Questions about this recall
Was The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of used irrigant. recalled?
Yes. FDA published recall Z-2555-2014 on September 3, 2014. Reported hazard: A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.
What should I do if I have this product?
Voluntary: Firm initiated · Letter