Cordis EMPIRA NC RX PTCA Dilatation Catheter
Published September 3, 2014 · Recall initiated: June 23, 2014
Summary
FDA records medical device Class II recall Z-2556-2014, published September 3, 2014 and initiated June 23, 2014. Recalling firm: Cordis Corporation; Status: Terminated; Reported quantity: 250,000 units; Lot: 63 Catalog numbers: 75R06200N, 75R06225N, 75R06250N, 75R06275N, 75R06300N, 75R06325N, 75R06350N, 75R06375N, 75R06400N, 75R10200N, 75R10225N, 75R10250N, 75R10275N, 75R10300N, 75R10325N, 75R10350N, 75R10375N, 75R10400N, 75R12200N, 75R12225N, 75R12250N, 75R12275N, 75R12300N, 75R12325N, 75R12350N, 75R12375N, 75R12400N, 75R15200N, 75R15225N, 75R15250N, 75R15275N, 75R15300N, 75R15325N, 75R15350N, 75R15375N, 75R15400N, 75R20200N, 75R20225N, 75R20250N, 75R20275N, 75R20300N, 75R20325N, 75R20350N, 75R20375N, 75R20400N, 75R25200N, 75R25225N, 75R25250N, 75R25275N, 75R25300N, 75R25325N, 75R25350N, 75R25375N, 75R25400N, 75R25200N, 75R25225N, 75R25250N, 75R25275N, 75R25300N, 75R25325N, 75R25350N, 75R25375N, and 75R25400N. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The balloon may exhibit resistance to the removal of the balloon cover. Weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover. This may lead to balloon burst, shaft separation or deflation difficulties, which could result in procedural delay, vessel trauma or surgery.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cordis Corporation
- Distributor
- Worldwide Distribution - USA including Puerto Rico and the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WY and Hawaii., and the countries of Argentina, Brazil, Canada, Chile, Colombia, Mexico, Panama and Venezuela.
- Lot
- 63 Catalog numbers: 75R06200N, 75R06225N, 75R06250N, 75R06275N, 75R06300N, 75R06325N, 75R06350N, 75R06375N, 75R06400N, 75R10200N, 75R10225N, 75R10250N, 75R10275N, 75R10300N, 75R10325N, 75R10350N, 75R10375N, 75R10400N, 75R12200N, 75R12225N, 75R12250N, 75R12275N, 75R12300N, 75R12325N, 75R12350N, 75R12375N, 75R12400N, 75R15200N, 75R15225N, 75R15250N, 75R15275N, 75R15300N, 75R15325N, 75R15350N, 75R15375N, 75R15400N, 75R20200N, 75R20225N, 75R20250N, 75R20275N, 75R20300N, 75R20325N, 75R20350N, 75R20375N, 75R20400N, 75R25200N, 75R25225N, 75R25250N, 75R25275N, 75R25300N, 75R25325N, 75R25350N, 75R25375N, 75R25400N, 75R25200N, 75R25225N, 75R25250N, 75R25275N, 75R25300N, 75R25325N, 75R25350N, 75R25375N, and 75R25400N.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASMART FLEX 6x150 BIL, 120cm The S.M.A.R.T.¿ Flex Biliary Stent System is indicated for use in the palliationNovember 24, 2021
- FDACordis Vista Brite Tip & Androit Guiding Catheters Product Usage: The guiding catheter is intended for use forApril 24, 2019
- FDACordis POWERFLEX PRO PTA Dilatation Catheter - Product Usage: Is intended to dilate stenoses in iliac,November 14, 2018
- FDACordis EMPIRA RX PTCA Dilatation CatheterSeptember 3, 2014
- FDAThe Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, andSeptember 3, 2014
- FDACURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUALSeptember 3, 2014
Questions about this recall
Was Cordis EMPIRA NC RX PTCA Dilatation Catheter recalled?
Yes. FDA published recall Z-2556-2014 on September 3, 2014. Reported hazard: The balloon may exhibit resistance to the removal of the balloon cover. Weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover. This may lead to balloon burst, shaft separation or deflation difficulties, which could result in procedural delay, vessel trauma or surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit