ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.

Published June 28, 2017 · Recall initiated: April 18, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Abbott Laboratories
Distributor
Worldwide Distribution - USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MONTENEGRO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UNITED KINGDOM, URUGUAY, VIETNAM
Lot
List number: 8K26-20, Lot Numbers: 01916E000, 00316F000, 01816G000, 00916I000, 07316I000, 00916L000 ; List Number: 8K26-25, Lot Numbers: 01816E000, 00216F000, 01716G000, 00816I000, 07216I000
Risk classification
Class II
Area (FDA)
device
Country
United States