StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.

Published August 5, 2020 · Recall initiated: June 5, 2020

Summary

FDA records medical device Class II recall Z-2707-2020, published August 5, 2020 and initiated June 5, 2020. Recalling firm: Medtronic Navigation, Inc.; Status: Terminated; Reported quantity: 2; Lot: StealthStation S7 Serial Number N06164844 with Polaris Spectra Camera Serial Number P7-01896; and StealthStation S7 Serial Number N07308697 with Polaris Spectra Camera Serial Number P7-18852. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Navigation, Inc.
Distributor
US Nationwide distribution including in the states of Alabama and Massachusetts.
Lot
StealthStation S7 Serial Number N06164844 with Polaris Spectra Camera Serial Number P7-01896; and StealthStation S7 Serial Number N07308697 with Polaris Spectra Camera Serial Number P7-18852.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. recalled?

Yes. FDA published recall Z-2707-2020 on August 5, 2020. Reported hazard: There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone