Transonic Flow Probe, 6 mm, Catalog Number HQC6FMV -- Product Usage: to measure flow intra-operatively.

Published September 9, 2015 · Recall initiated: July 27, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Transonic Systems Inc
Distributor
Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
Lot
Serial Number: HQC6FMV2239
Risk classification
Class II
Area (FDA)
device
Country
United States