Transonic Flow Probe, 6 mm, Catalog Number HQC6FMV -- Product Usage: to measure flow intra-operatively.
Published September 9, 2015 · Recall initiated: July 27, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Transonic Systems Inc
- Distributor
- Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
- Lot
- Serial Number: HQC6FMV2239
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States