Back to search
FDAFDA · Food and drugs·Recall number Z-2722-2015

Transonic Flow Probe, 6 mm, Catalog Number HQC6FMV -- Product Usage: to measure flow intra-operatively.

Published September 9, 2015 · Recall initiated: July 27, 2015

Summary

FDA records medical device Class II recall Z-2722-2015, published September 9, 2015 and initiated July 27, 2015. Recalling firm: Transonic Systems Inc; Status: Terminated; Reported quantity: Domestic: 1 unit; Lot: Serial Number: HQC6FMV2239. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Transonic Systems Inc
Distributor
Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
Lot
Serial Number: HQC6FMV2239
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Transonic Flow Probe, 6 mm, Catalog Number HQC6FMV -- Product Usage: to measure flow intra-operatively. recalled?

Yes. FDA published recall Z-2722-2015 on September 9, 2015. Reported hazard: Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

What should I do if I have this product?

Voluntary: Firm initiated · Letter