Transonic Flow Probe, 2 mm, Catalog Number HQD2FMC -- Product Usage: to measure flow intra-operatively.

Published September 9, 2015 · Recall initiated: July 27, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Transonic Systems Inc
Distributor
Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
Lot
Serial Numbers: HQD2FMC2797, HQD2FMC2873, HQD2FMC2874, HQD2FMC2875, HQD2FMC2876, HQD2FMC2877, HQD2FMC2878, HQD2FMC2879, HQD2FMC2880, HQD2FMC2881, HQD2FMC2882, HQD2FMC2883, HQD2FMC2884, HQD2FMC2885, HQD2FMC2886, HQD2FMC2888, HQD2FMC2889, HQD2FMC2890, HQD2FMC2908
Risk classification
Class II
Area (FDA)
device
Country
United States