Transonic Flow Probe, 8 mm, Catalog Number HQC8FMV -- Product Usage: to measure flow intra-operatively.
Published September 9, 2015 · Recall initiated: July 27, 2015
Summary
FDA records medical device Class II recall Z-2723-2015, published September 9, 2015 and initiated July 27, 2015. Recalling firm: Transonic Systems Inc; Status: Terminated; Reported quantity: Domestic: 5 units; Lot: Serial Numbers: HQC8FMV2142, HQC8FMV2143, HQC8FMV2144, HQC8FMV2145, HQC8FMV2146. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Transonic Systems Inc
- Distributor
- Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
- Lot
- Serial Numbers: HQC8FMV2142, HQC8FMV2143, HQC8FMV2144, HQC8FMV2145, HQC8FMV2146
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATransonic ADT1018 Flow QC Clear Advantage Tubing, Catalog No. ADT1018, ADT1018-40 The device is indicated forMay 16, 2018
- FDATransonic Flow Probe, 2 mm, Catalog Number HQC2FMC Product Usage: to measure flow intra-operatively.September 9, 2015
- FDATransonic Flow Probe, 2 mm, Catalog Number HQD2FMC -- Product Usage: to measure flow intra-operatively.September 9, 2015
- FDATransonic Flow Probe, 6 mm, Catalog Number HQC6FMV -- Product Usage: to measure flow intra-operatively.September 9, 2015
- FDAMagic3 Intermittent Catheter with Sure Grip Hydrophilic Coated Silicone Coude tip, Product Code 50614. TheSeptember 9, 2015
- FDAUROLOGY PACK 5/CS Used by medical staff to perform surgical procedures and patient care procedures such asSeptember 9, 2015
Questions about this recall
Was Transonic Flow Probe, 8 mm, Catalog Number HQC8FMV -- Product Usage: to measure flow intra-operatively. recalled?
Yes. FDA published recall Z-2723-2015 on September 9, 2015. Reported hazard: Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
What should I do if I have this product?
Voluntary: Firm initiated · Letter