Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling Catheter with 3M(TM) SteriStrip(TM), (1) Pump Tubing Set, (1) Drainage Bag. Medtronic Navigation, Inc. Louisville, CO 80027

Published August 2, 2017 · Recall initiated: June 7, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Medtronic Navigation, Inc.
Distributor
US Distribution to the states of : CA, GA, MN, FL, NY, IL, TX
Lot
Lot Numbers: 211158059, 211560973, 211959879, 12473141, 212707035, 211158151, 211594328, 211977586, 212503634, 212745660, 211158152, 211599898, 211987873, 212504463, 212796559, 211181212, 211628257, 212127654, 212549456, 212912501, 211181547,211908695, 212141557, 212553428, 211184941, 211916062, 212148368, 212632884,211555011, 211926889, 212434500, 212647830, 211555017, 211950736, 212438405, 212701679