FDAFDA · Food and drugs·Recall number Z-2746-2017
Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958.
Published July 26, 2017 · Recall initiated: May 10, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated Elevate Inserter with the StealthAiR Spine Frame during spine surgical procedures.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Medtronic Navigation, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Brazil, Germany, Switzerland and Great Britain.
- Lot
- Kits that include affected 9731958 INSTALL CD: INST SET 9734833 UNIVERSAL NAVLOCK, INST SET 9734834 CAPSTONE TRIALS, INST SET 9734835 CLYDESDALE TRIALS, INST SET 9734836 TLIF/DLIF INSERTER, INST SET 9735283 SOLERA 5.5/6.0 DRIVERS, INST SET 9735284 SOLERA NONCAN TAPS, INST SET 9735285 SOLERA ILIAC TAPS, TAPS KIT 9735465 SOLERA AWL-TIP TAPS, REF SET 9735500 SPINE REFRENCING, UPGRD KIT 9735522 SPINE REVISION, INSTR SET 9735607 VERTEX SELECT
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?