FDAFDA · Food and drugs·Recall number Z-2808-2026
EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622
Published August 5, 2026 · Recall initiated: June 24, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Medtronic Perfusion Systems
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- UDI-DI: 20763000135670, Lot Numbers: 0232601501 0232601510 0232601549 0232601550 0232602178 0232602290 0232602302 0232910952 0232910953 0232910954 0232910955 0233244075 0233244301 0233244531 0233286362 0233286374 0233286390 0233409034 0233416071 0233528731
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.