HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU)
Published September 9, 2026 · Recall initiated: July 31, 2026
Summary
FDA records medical device Class II recall Z-3064-2026, published September 9, 2026 and initiated July 31, 2026. Recalling firm: Johnson & Johnson Surgical Vision, Inc.; Status: Ongoing; Reported quantity: 111,478; Lot: REF/UDI-DI/Lot: 10300012/5050474755093/UA32232, UA32250, UA32268, UA32290, UA32292, UA32303, UA32371, UP31203, UP31205; 10300016/5050474755116/UA32447, UA32448, UA32449, UB33008, UB33026, UB33082, UB33083, UB33175; 10301012/5050474755109/UB33014, UB33041, UB33057, UB33060, UB33100, UB33190; 10301016/5050474755123/UB33084, UB33194. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Johnson & Johnson Surgical Vision, Inc.
- Distributor
- Worldwide - US Nationwide distribution including in the states of PA, AL, AZ, AR, CA, FL, IL, IN, IA, KY, LA, ME, MI, MN, MS, NJ, NY, NC, OH, OR, TN, TX, WA, WV, HI, CO, MD, VA, WI, ID, NH, SD, OK, MA, NM, DE, GA, SC, CT, NV, MO, RI, KS, ND and the countries of CA, BE, CH, DE, DK, ES, FR GB, HU, IS, IT, NL, NO, PT, SE.
- Lot
- REF/UDI-DI/Lot: 10300012/5050474755093/UA32232, UA32250, UA32268, UA32290, UA32292, UA32303, UA32371, UP31203, UP31205; 10300016/5050474755116/UA32447, UA32448, UA32449, UB33008, UB33026, UB33082, UB33083, UB33175; 10301012/5050474755109/UB33014, UB33041, UB33057, UB33060, UB33100, UB33190; 10301016/5050474755123/UB33084, UB33194
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, CatalogSeptember 9, 2026
- FDABaxter Novum IQ Syringe Pump, Model Number 40800BAXUSSeptember 9, 2026
- FDAMedline Convenience Kits: 1) ORTHO ASPIRATION KIT, Model Number: DYKM1636A; 2) SHOULD BLOCK KIT, Model Number:September 9, 2026
- FDABaxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSLSeptember 9, 2026
- FDAJohnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21LMay 8, 2024
- FDATECNIS Toric II OptiBlue IOL Models ZCWJanuary 24, 2024
Questions about this recall
Was HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU) recalled?
Yes. FDA published recall Z-3064-2026 on September 9, 2026. Reported hazard: Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
What should I do if I have this product?
Voluntary: Firm initiated · Letter