TECNIS Toric II OptiBlue IOL Models ZCW
Published January 24, 2024 · Recall initiated: December 12, 2023
Summary
FDA records medical device Class II recall Z-0735-2024, published January 24, 2024 and initiated December 12, 2023. Recalling firm: Johnson & Johnson Surgical Vision, Inc.; Status: Ongoing; Reported quantity: 36; Lot: Model Numbers: ZCW375 UDI code: N/A (not distributed in the U.S.) Serial Numbers: 3473902107 3473912107 3473922107 3473932107 3473942107 3473952107 3473962107 3473972107 3474002107 3474012107 3474022107 3474072107 3474082107 3474092107 3474102107 3474112107 3474122107 3474132107 3474152107 3474162107 3474172107 3474182107 3474192107 3474202107 3474222107 3474232107 3474242107 3474252107 3474262107 3474282107 3474292107 3474312107 3474322107 3474332107 3474342107 3474352107. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle between Low Power Meridian (LPM) and toric cylinder axis marks.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Johnson & Johnson Surgical Vision, Inc.
- Distributor
- International distribution to the country of Japan.
- Lot
- Model Numbers: ZCW375 UDI code: N/A (not distributed in the U.S.) Serial Numbers: 3473902107 3473912107 3473922107 3473932107 3473942107 3473952107 3473962107 3473972107 3474002107 3474012107 3474022107 3474072107 3474082107 3474092107 3474102107 3474112107 3474122107 3474132107 3474152107 3474162107 3474172107 3474182107 3474192107 3474202107 3474222107 3474232107 3474242107 3474252107 3474262107 3474282107 3474292107 3474312107 3474322107 3474332107 3474342107 3474352107
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO,September 9, 2026
- FDAJohnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21LMay 8, 2024
- FDAFreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 OperatingJanuary 24, 2024
- FDAPosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill),January 24, 2024
- FDAVACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate, Item #454297, 13x75 grey cap-black ring, non-ridged.January 24, 2024
- FDAV. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18January 24, 2024
Questions about this recall
Was TECNIS Toric II OptiBlue IOL Models ZCW recalled?
Yes. FDA published recall Z-0735-2024 on January 24, 2024. Reported hazard: Due to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle between Low Power Meridian (LPM) and toric cylinder axis marks.
What should I do if I have this product?
Voluntary: Firm initiated · Letter