Spine Referencing Instrumentation, Spinous Process Clamp, Tall When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.
Published September 27, 2017 · Recall initiated: June 29, 2017
Summary
FDA records medical device Class II recall Z-3139-2017, published September 27, 2017 and initiated June 29, 2017. Recalling firm: Medtronic Navigation, Inc.; Status: Terminated; Reported quantity: 2,051 units total; Lot: P/N 9734715, 9734715K. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Navigation, Inc.
- Distributor
- Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, AD, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, & WY. Foreign: Albania, Australia, Bangladesh, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Kenya, Latvia, Luxembourg, Malaysia, Martinique, Netherlands, New Zealand, Philippines, Poland, Portugal, Romania, Russia, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, & Vietnam.
- Lot
- P/N 9734715, 9734715K
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801075, Sterile EO, 5/Tray 12 Pk., RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., RxonlyOctober 28, 2026
- FDAK-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R)September 27, 2017
- FDASpine Referencing Instrumentation, Double Spinous Process Clamp, Tall When used with a MedtronicSeptember 27, 2017
Questions about this recall
Was Spine Referencing Instrumentation, Spinous Process Clamp, Tall When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp. recalled?
Yes. FDA published recall Z-3139-2017 on September 27, 2017. Reported hazard: Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit