OMNIFIT SERIES I ACET. INSERT 10® 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Published September 23, 2026 · Recall initiated: July 30, 2026
Summary
FDA records medical device Class II recall Z-3196-2026, published September 23, 2026 and initiated July 30, 2026. Recalling firm: Howmedica Osteonics Corp.; Status: Ongoing; Reported quantity: 12,156 US; Lot: UDI-DI 07613327037883 Lot 13MFT. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Howmedica Osteonics Corp.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Lot
- UDI-DI 07613327037883 Lot 13MFT
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.October 21, 2026
- FDATriathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.October 21, 2026
- FDATS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.October 21, 2026
- FDATS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.October 21, 2026
- FDAAXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufacturedSeptember 23, 2026
- FDAALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318September 23, 2026
Questions about this recall
Was OMNIFIT SERIES I ACET. INSERT 10® 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured from UHMWPE (polyethylene) recalled?
Yes. FDA published recall Z-3196-2026 on September 23, 2026. Reported hazard: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
What should I do if I have this product?
Voluntary: Firm initiated · Letter