AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Published September 23, 2026 · Recall initiated: July 30, 2026

Summary

FDA records medical device Class II recall Z-3195-2026, published September 23, 2026 and initiated July 30, 2026. Recalling firm: Howmedica Osteonics Corp.; Status: Ongoing; Reported quantity: 12,156 US; Lot: No UDI-DI Lot P22E159. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Howmedica Osteonics Corp.
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Lot
No UDI-DI Lot P22E159
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene) recalled?

Yes. FDA published recall Z-3195-2026 on September 23, 2026. Reported hazard: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

What should I do if I have this product?

Voluntary: Firm initiated · Letter