prodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5
Published October 2, 2024 · Recall initiated: July 18, 2024
Summary
FDA records medical device Class II recall Z-3213-2024, published October 2, 2024 and initiated July 18, 2024. Recalling firm: Centinel Spine, Inc.; Status: Ongoing; Reported quantity: 6 units; Lot: Lot Code: UDI-DI 00843193113856; Lot 2024-0482. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Centinel Spine, Inc.
- Distributor
- US Nationwide distribution.
- Lot
- Lot Code: UDI-DI 00843193113856; Lot 2024-0482
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProdisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.May 13, 2026
- FDAProdisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.May 13, 2026
- FDAprodisc C SK, Total cervical disc replacement, medical device, Model/Catalog Number: PDSM5October 2, 2024
- FDAGE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.October 2, 2024
- FDAAVS Anchor-C Cervical Cage System Instructions For Use (IFU)- eIFU (revision 5)-Indicated for anteriorOctober 2, 2024
- FDAACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON JuniperOctober 2, 2024
Questions about this recall
Was prodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5 recalled?
Yes. FDA published recall Z-3213-2024 on October 2, 2024. Reported hazard: During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone