GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.
Published October 2, 2024 · Recall initiated: August 5, 2024
Summary
FDA records medical device Class II recall Z-3214-2024, published October 2, 2024 and initiated August 5, 2024. Recalling firm: GE Healthcare; Status: Ongoing; Reported quantity: 2138 units; Lot: All versions of Centricity PACS: V2, UDI/DI Not Applicable; V3, UDI/DI Not Applicable; V4, UDI/DI 00840682124430; V6, UDI/DI 00840682104807; V7, UDI/DI 00840682145572 installed on the following systems: CARTI02312611, CARTI22214087T, 501268SSASW, 501268SSATSW, 928757VPASW, 928757VPBSW, 928757VPTSW, 602344PVSW2, 602344REPSITEB, MARICOPAC02308029, 602344TIMSVM, 480323PVSW, 661633PVSW, 805395PVSW, 805395PVTSW, V5SPORTS01804913, 310349PACSSWPV, UCLAMC2259090, 310349PACSSWBPV, UCLAMC2259063, 831624PVASW, UHSD02184496, 916781CPACSSRWT, PALMDALEM01810792, PALMDALEM16023955, UHSD16084343, PALMDALEM20508297, PALMDALEM21178954, MHL02186470, 909788HCAPACS1, 916781CPACSCKS, 916781CPACSBCKS, 650696PACSEESW, MPHS2259091, GSHCACPSW, GSH320037250, VSMY01820423, 209532SESW, SONORAREG19607823, RANCHO02255039, RANCHOPISU2, RANCHOPTESTISU, To be Provided , 805370HCALR, VMPACS69, LOSROBLES19666790, SOUTHWEST01815383, MCANPACSW, MCASPACSW, CENPACSW, 303329HCAPSWS, 303584HCAPVSW, 303584REPBSW, 303584HCAPVSWTESTIMS, SRMCPACSW, 719584VPSW, PKV2259020, 719584VPBSW, NSMCPACSW, 970476PVSW, VVMC15111181, 302674PACSSW, 302674PACSBSW, 302736SAEATSW, JFKSW, JFKR19980220B, INFINITYW01452121, 727462PVSW, 727462PVBSW, MPMH2259089T, 772461HCAPACSW, 772461HCAPACSW1, 772461HCAPACSWT, FWMCCPTSW, 352333PACSSW, NFRH19663978B, 904202BMCDSW, 904202BMCSSW, MJPACSW, MJPACTSW, 904202BMCDTSW, BIHT20915070-IMSTest2, BIHT20915071-IMSTEST4x, 407518COREPACS, HH1001398036, 407518CPREPB, 863680PVSW, KRMCOREPACSSW, KRMC19478073, 407888HCASWSS, 615344HCASW, PCMC420084442, 561694CAPACSSW, WFLHIV01, WESTFLORI19661467, 772335PACSSW, 772335PVSW, 941637PVSW, 941637TIMS90, 850875PVSW2, 904819PVSW, 904819PV2SW, 904819PVTSW, 727519PVBSW, 850325PACSCORE, CRMC820084433, 888663CPBSW, USFIMSSSASW, 727519PVSW, VSMAE01817090, NGDC01893569, 888663CPTSW, 888663CPB2SW, USFIMSSSABSW, USFIMSSSATSW1, USFIMSSSATSW2, 727519PVMSW, 727519PVTSW, VAMAF02345966, VSMI18957442DCC, 407518SCPACSHC, 850325HCASWOPT, WFLHIVT01, PCMC42259036, 352333PACSTSW, VSMAM02004646, WELLINGTON01810730, UHSD02289750, WELLINGTON19678279T, WELLINGTON20036598, TRMPACSW, TRMNC20021605, 100115-8-Centricity PACS-21222171, TRMTESTPACS, Doctors Ho22222417, 407772CORESW, 404605PVBSW, 404851PVSW, 404851PVTSW, 404851PVBSW, 404605PVTSW, 407772COREBSW, 407772CORETSW, 706651HCACORE, 912466PVSW, SGHS2259039, 912276PVSW, 912276PVTSW, 706278PACSSWPV, 706278PACSSWBPV, 706278PACSSWTPV, 404501PACSSWPV, 404501PACSSWBPV, 404501PACSSWATPV, 404501PACSSWBTPV, 706754PACSSW, 404605PVSW, 478633PACSSWPV, MCG2259011, MCG20084728, VSMQL01987331, 404325PACSSWPV, 404325PACSSWBPV, 404325PACSTSW, 406238PVSW, VSMZ01812192, 641422PVSW, PROSPECT2259057, PROSPECT2259084, PROSPECT22128566, 319272SCPSWULP, 319272CPDRSW, 319272CPDRBSW, 319272SCPTSWULP, 208COREPVSW, EID19660790, 208CORETIMS, 208743PVSW, 208743PVTSW, 847NWCPVSW, 847NWCPVSW2, 773880PVSW, 795079PACSEESW1, 795079PACSEESW2, 312926PACSSWPV, 312996PVSW2, JSHPACSSW, 312996PVSW1, 312996PVSWT1, 312996PVSWT2, 795079PACST, JHSJHOCC02259030, 773880PVBSW, 773880PVTSW, JSHPACSTSW, 217876PVSW, 217876PVBSW, 815285PACUV100, 630963PVSW, ADVPACSTSW1, 847437PVSW, 847437DCPVSW, 847437SAPVTSW, 847990PVSW, S046PACSSWPV, S801PACSSWPV, S046PACSSWTPV, ADVPACSSW, ADVPACSBSW, ADVPACSTSW, 815398CPACSSW, 219325PVSW1, 219325PVBSW1, 219325PVTSW1, 765983PVSW, 765983PVSW2, RHHCS02253964T, 812232PACSPV, 812232PACSTPV, 812232PACS2PV, THRH20084364, FRNKHAPACSPV, FRNKHAPACSTPV, FRANKFORT19633100, RRMCPACSW, RGMCPACSTW, RRMC120080697, VTP01810617T, VHT02334235P, VTP19617532, 985867CORESW, 985230PVSW, NOMC2259032, CFO18227240, 985230PVSWS, 504903CPACSSW, 504988SSASW, 504903CPACSTSW, 504988SSATSW, MCOLANO20039031, HTMC20085961, 504903CPACSIMS, 617638SSASW, 617638SSBSW, 617638SSATSW, 781979PVSW, 781979PVBSW, 781979PVTSW, 413447BCPVSW, 413447HCPVSW, 413447PVSW, 413748PVSW, 413748PVTSW, 413748PVBSW, 301698PVSW, 301698PVTSW, NORTHE02315652, NORTHE02370307T, 207795PVSWC3, 207795PVSWC1, 207795PVSWC2, 517353PACSEESW, 517353PACSEETSW, 906387VPSW. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare
- Distributor
- Worldwide distribution.
- Lot
- All versions of Centricity PACS: V2, UDI/DI Not Applicable; V3, UDI/DI Not Applicable; V4, UDI/DI 00840682124430; V6, UDI/DI 00840682104807; V7, UDI/DI 00840682145572 installed on the following systems: CARTI02312611, CARTI22214087T, 501268SSASW, 501268SSATSW, 928757VPASW, 928757VPBSW, 928757VPTSW, 602344PVSW2, 602344REPSITEB, MARICOPAC02308029, 602344TIMSVM, 480323PVSW, 661633PVSW, 805395PVSW, 805395PVTSW, V5SPORTS01804913, 310349PACSSWPV, UCLAMC2259090, 310349PACSSWBPV, UCLAMC2259063, 831624PVASW, UHSD02184496, 916781CPACSSRWT, PALMDALEM01810792, PALMDALEM16023955, UHSD16084343, PALMDALEM20508297, PALMDALEM21178954, MHL02186470, 909788HCAPACS1, 916781CPACSCKS, 916781CPACSBCKS, 650696PACSEESW, MPHS2259091, GSHCACPSW, GSH320037250, VSMY01820423, 209532SESW, SONORAREG19607823, RANCHO02255039, RANCHOPISU2, RANCHOPTESTISU, To be Provided , 805370HCALR, VMPACS69, LOSROBLES19666790, SOUTHWEST01815383, MCANPACSW, MCASPACSW, CENPACSW, 303329HCAPSWS, 303584HCAPVSW, 303584REPBSW, 303584HCAPVSWTESTIMS, SRMCPACSW, 719584VPSW, PKV2259020, 719584VPBSW, NSMCPACSW, 970476PVSW, VVMC15111181, 302674PACSSW, 302674PACSBSW, 302736SAEATSW, JFKSW, JFKR19980220B, INFINITYW01452121, 727462PVSW, 727462PVBSW, MPMH2259089T, 772461HCAPACSW, 772461HCAPACSW1, 772461HCAPACSWT, FWMCCPTSW, 352333PACSSW, NFRH19663978B, 904202BMCDSW, 904202BMCSSW, MJPACSW, MJPACTSW, 904202BMCDTSW, BIHT20915070-IMSTest2, BIHT20915071-IMSTEST4x, 407518COREPACS, HH1001398036, 407518CPREPB, 863680PVSW, KRMCOREPACSSW, KRMC19478073, 407888HCASWSS, 615344HCASW, PCMC420084442, 561694CAPACSSW, WFLHIV01, WESTFLORI19661467, 772335PACSSW, 772335PVSW, 941637PVSW, 941637TIMS90, 850875PVSW2, 904819PVSW, 904819PV2SW, 904819PVTSW, 727519PVBSW, 850325PACSCORE, CRMC820084433, 888663CPBSW, USFIMSSSASW, 727519PVSW, VSMAE01817090, NGDC01893569, 888663CPTSW, 888663CPB2SW, USFIMSSSABSW, USFIMSSSATSW1, USFIMSSSATSW2, 727519PVMSW, 727519PVTSW, VAMAF02345966, VSMI18957442DCC, 407518SCPACSHC, 850325HCASWOPT, WFLHIVT01, PCMC42259036, 352333PACSTSW, VSMAM02004646, WELLINGTON01810730, UHSD02289750, WELLINGTON19678279T, WELLINGTON20036598, TRMPACSW, TRMNC20021605, 100115-8-Centricity PACS-21222171, TRMTESTPACS, Doctors Ho22222417, 407772CORESW, 404605PVBSW, 404851PVSW, 404851PVTSW, 404851PVBSW, 404605PVTSW, 407772COREBSW, 407772CORETSW, 706651HCACORE, 912466PVSW, SGHS2259039, 912276PVSW, 912276PVTSW, 706278PACSSWPV, 706278PACSSWBPV, 706278PACSSWTPV, 404501PACSSWPV, 404501PACSSWBPV, 404501PACSSWATPV, 404501PACSSWBTPV, 706754PACSSW, 404605PVSW, 478633PACSSWPV, MCG2259011, MCG20084728, VSMQL01987331, 404325PACSSWPV, 404325PACSSWBPV, 404325PACSTSW, 406238PVSW, VSMZ01812192, 641422PVSW, PROSPECT2259057, PROSPECT2259084, PROSPECT22128566, 319272SCPSWULP, 319272CPDRSW, 319272CPDRBSW, 319272SCPTSWULP, 208COREPVSW, EID19660790, 208CORETIMS, 208743PVSW, 208743PVTSW, 847NWCPVSW, 847NWCPVSW2, 773880PVSW, 795079PACSEESW1, 795079PACSEESW2, 312926PACSSWPV, 312996PVSW2, JSHPACSSW, 312996PVSW1, 312996PVSWT1, 312996PVSWT2, 795079PACST, JHSJHOCC02259030, 773880PVBSW, 773880PVTSW, JSHPACSTSW, 217876PVSW, 217876PVBSW, 815285PACUV100, 630963PVSW, ADVPACSTSW1, 847437PVSW, 847437DCPVSW, 847437SAPVTSW, 847990PVSW, S046PACSSWPV, S801PACSSWPV, S046PACSSWTPV, ADVPACSSW, ADVPACSBSW, ADVPACSTSW, 815398CPACSSW, 219325PVSW1, 219325PVBSW1, 219325PVTSW1, 765983PVSW, 765983PVSW2, RHHCS02253964T, 812232PACSPV, 812232PACSTPV, 812232PACS2PV, THRH20084364, FRNKHAPACSPV, FRNKHAPACSTPV, FRANKFORT19633100, RRMCPACSW, RGMCPACSTW, RRMC120080697, VTP01810617T, VHT02334235P, VTP19617532, 985867CORESW, 985230PVSW, NOMC2259032, CFO18227240, 985230PVSWS, 504903CPACSSW, 504988SSASW, 504903CPACSTSW, 504988SSATSW, MCOLANO20039031, HTMC20085961, 504903CPACSIMS, 617638SSASW, 617638SSBSW, 617638SSATSW, 781979PVSW, 781979PVBSW, 781979PVTSW, 413447BCPVSW, 413447HCPVSW, 413447PVSW, 413748PVSW, 413748PVTSW, 413748PVBSW, 301698PVSW, 301698PVTSW, NORTHE02315652, NORTHE02370307T, 207795PVSWC3, 207795PVSWC1, 207795PVSWC2, 517353PACSEESW, 517353PACSEETSW, 906387VPSW
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAWorkflow Core Services V8, intended to provide set of services for storage and retrieval of patient data,September 23, 2026
- FDACentricity Enterprise Archive V2, V3, and V4, software product for receiving, archiving, and sendingSeptember 23, 2026
- FDACentricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review,September 23, 2026
- FDACore Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC)September 23, 2026
- FDAprodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5October 2, 2024
- FDAACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON JuniperOctober 2, 2024
Questions about this recall
Was GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system. recalled?
Yes. FDA published recall Z-3214-2024 on October 2, 2024. Reported hazard: GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
What should I do if I have this product?
Voluntary: Firm initiated · Letter