curaplex Epi Safe Kit, 8600-01101, contains:" 1ml Vial of Epinephrine " 1 Epi-Safe Syringe " 1 Safety Needle " 2 Alcohol Prep Pads " 1 Adhesive Dressing " 1 Insert, Rx Only. Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016

Publicado el 30 de enero de 2019 · Inicio del retiro: 2 de noviembre de 2018

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Bound Tree Medical, LLC
Distribuidor
Nationwide in the USA
Lote
all lot numbers
Clasificación de riesgo
Class II
Área (FDA)
drug