Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0013-2017

Multix Fusion Stationary X-Ray System Product Usage: The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill)applications.

Publicado el 12 de octubre de 2016 · Inicio del retiro: 1 de agosto de 2016

Resumen

FDA registra el retiro Z-0013-2017 de dispositivos médicos de Clase II, publicado el 12 de octubre de 2016 e iniciado el 1 de agosto de 2016. Empresa responsable: Siemens Medical Solutions USA, Inc; Estado: Finalizado; Cantidad reportada: 37 Systems; Lote: Model Number: 10762470, 10893300, 10746665 Serial Numbers: 24644,24613,24615,40328,24611,24612,24614,40326,40322,40316,24599,24600,40404,24634,2464124590,40402,1314,40312,24591,24592,40327,40323,24656,24659,40335,1315,40332,40403,24605,40325,40313,24645,24610,24580,24648,24581. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Siemens became aware of a potential issue with the patient table for Ysio Max and Multix Fusion systems. There is an unlikely risk that an internal part of the table may detach causing the table top to tilt to the left side which could result in an injury to a patient on the table.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Siemens Medical Solutions USA, Inc
Distribuidor
US Nationwide Distribution
Lote
Model Number: 10762470, 10893300, 10746665 Serial Numbers: 24644,24613,24615,40328,24611,24612,24614,40326,40322,40316,24599,24600,40404,24634,2464124590,40402,1314,40312,24591,24592,40327,40323,24656,24659,40335,1315,40332,40403,24605,40325,40313,24645,24610,24580,24648,24581
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Multix Fusion Stationary X-Ray System Product Usage: The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill)applications. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0013-2017 el 12 de octubre de 2016. Peligro reportado: Siemens became aware of a potential issue with the patient table for Ysio Max and Multix Fusion systems. There is an unlikely risk that an internal part of the table may detach causing the table top to tilt to the left side which could result in an injury to a patient on the table.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter