Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Publicado el 16 de octubre de 2024 · Inicio del retiro: 18 de abril de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Beckman Coulter, Inc.
- Distribuidor
- Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
- Lote
- UDI-DI: 15099590230623; U.S. serial numbers: 300284, 300286, 300294, 300266, 300291, 300223, 300260, 300285, 300292, 300195, 300227, 300228, 300205, 300207, 300206, 300182, 300192, and 300234. OUS serial numbers have been requested.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States