Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; List Number 16026-01 Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural
Publicado el 31 de octubre de 2012 · Inicio del retiro: 29 de agosto de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a delay in therapy or result in over delivery or under delivery of medication if the user does not confirm the programmed values on the pump's confirmation screen prior to starting the infusion.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Hospira Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Canada
- Lote
- List Number 16026-01, all serial numbers
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States