Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0078-2025

Avance. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.

Publicado el 30 de octubre de 2024 · Inicio del retiro: 12 de julio de 2024

Resumen

FDA registra el retiro Z-0078-2025 de dispositivos médicos de Clase I, publicado el 30 de octubre de 2024 e iniciado el 12 de julio de 2024. Empresa responsable: Datex-Ohmeda, Inc.; Estado: En curso; Cantidad reportada: 9,349 units; Lote: GTIN: 00840682102322. Model Number: 1009-9002-000. All Serial Numbers. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Datex-Ohmeda, Inc.
Distribuidor
Worldwide distribution - US Nationwide and the countries of Aland Islands, ALBANIA, Algeria, Argentina, Australia, Austria, Azerbaijan, BAHAMAS, Bahrain, Bangladesh, BARBADOS, BELARUS, Belgium, Benin, Bermuda, Bolivia, BONAIRE, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Cameroon, Canada, Chile, CHINA, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, ETHIOPIA, Falkland Islands, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gilbrator, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, ICELAND, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Jamaica, JAPAN, Jordan, Kazakhstan, KENYA, Korea (Republic Of), Korea, Republic of, Kuwait, Laos, LATVIA, Lebanon, Libya, LITHUANIA, Macedonia, Malaysia, Maldives, Malta, MAURITIUS, Mexico, Moldova, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, PAKISTAN, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, PUERTO RICO, Qatar, REUNION, ROMANIA, Russia, San Marino, Saudi Arabia, Senegal, Serbia, SINGAPORE, Slovakia, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, THAILAND, Trinidad and Tobago, Tunisia, Turkey, T®rkiye, Uganda, UKRAINE, United Arab Emirates, United Kingdom, Uruguay, UZBEKISTAN, Venezuela, Vietnam, Yemen.
Lote
GTIN: 00840682102322. Model Number: 1009-9002-000. All Serial Numbers. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Avance. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0078-2025 el 30 de octubre de 2024. Peligro reportado: Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter