Patient Cart used in conjunction with the da Vinci S surgical and SI System. Intuitive Surgical 1266 Kifer Road Sunnyvale, CA 94086 The Patient Side Cart (PSC) is a robotic platform used for da Vinci Surgery. It is the operative component for both da Vinci systems and its primary function is to support he instrument arms and camera arm. It uses remote center technology. The remote center is a fixed point in space around which the Patient Cart arms move. Remote center technology enables the system to maneuver instruments and endoscopes in the surgical site while exerting minimal force on the patient body wall. The Patient Cart works in the sterile field, assisting the Surgeon Console operator by exchanging instruments and endoscopes, and by performing other patient-side activities. To help ensure patient safety, the action of the Patient Cart operator take precedence over actions of the Surgeon Console operator.

Publicado el 13 de noviembre de 2013 · Inicio del retiro: 11 de diciembre de 1930

Resumen

FDA registra el retiro Z-0139-2014 de dispositivos médicos de Clase II, publicado el 13 de noviembre de 2013 e iniciado el 11 de diciembre de 1930. Empresa responsable: Intuitive Surgical, Inc.; Estado: Finalizado; Cantidad reportada: 1754 patient carts; Lote: Model IS2000 and IS 3000. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Increased number of complaints regarding fluid entering the patient card base of the da Vinci S and SI systems.

Qué hacer

Voluntary: Firm initiated · Visit

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Intuitive Surgical, Inc.
Distribuidor
Worldwide Distribution - USA including Puerto Rico and the countries of Australia, Austria, Argentina, Belgium, Bulgaria, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Japan, Kuwait, Mexico, Netherlands, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uruguay and Venezuela.
Lote
Model IS2000 and IS 3000
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Patient Cart used in conjunction with the da Vinci S surgical and SI System. Intuitive Surgical 1266 Kifer Road Sunnyvale, CA 94086 The Patient Side Cart (PSC) is a robotic platform used for da Vinci Surgery. It is the operative component for both da Vinci systems and its primary function is to support he instrument arms and camera arm. It uses remote center technology. The remote center is a fixed point in space around which the Patient Cart arms move. Remote center technology enables the system to maneuver instruments and endoscopes in the surgical site while exerting minimal force on the patient body wall. The Patient Cart works in the sterile field, assisting the Surgeon Console operator by exchanging instruments and endoscopes, and by performing other patient-side activities. To help ensure patient safety, the action of the Patient Cart operator take precedence over actions of the Surgeon Console operator. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0139-2014 el 13 de noviembre de 2013. Peligro reportado: Increased number of complaints regarding fluid entering the patient card base of the da Vinci S and SI systems.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Visit