Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0164-2015

Siemens Calibrator E; SMN 10309079 (2 pack), SMN 10321075 (6 pack), SMN 10335532 (6 pack Ref)

Publicado el 12 de noviembre de 2014 · Inicio del retiro: 26 de septiembre de 2014

Resumen

FDA registra el retiro Z-0164-2015 de dispositivos médicos de Clase II, publicado el 12 de noviembre de 2014 e iniciado el 26 de septiembre de 2014. Empresa responsable: Siemens Healthcare Diagnostics, Inc; Estado: Finalizado; Cantidad reportada: Total = 19,352 (13,881 - 2 pack, 4878 - 6 pack, 593 - 6 pack REF); Lote: 26284A38, 26900A38, 27991A38, 28192A38, 29571A38, 31162A38, 32488A38, 35243AB39, 35295AB39, 36091AB39, 39464AB39, 40245AB39, 41317AB39, 26285A38, 27348A38, 29169A38, 30710A38, 31261A38, 32687A38, 35259AB39, 35359AB39, 36289AB39, 39031AB39, 41124AB39, 27236A38, 35051A38, 35839AB39, 40141AB39, 41702AB39. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Positive bias for Calibrator E lot kits ending in 38 and 39 with the ADVIA Centaur Systems Cortisol and Progesterone assays compared to the respective master curves.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Siemens Healthcare Diagnostics, Inc
Distribuidor
Nationwide including Puerto Rico. Worldwide: Afghanistan, Armenia, Austria, Bahrain, Belgium, Bulgaria, Burkina Faso, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Georgia, Germany, Great Britain, Greece, Hungary, Iran, Iraq, Ireland, Italy, Jordan, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malta, Morocco, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Unit.Arab Emir., Vatican, Yemen, AR, AU, BD, BR, CA , CH, CL, CN, CO, DO, EC, EG, GT, HK, ID, IE, IL, IN, JP, KR, LY, MM, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VN, ZA
Lote
26284A38, 26900A38, 27991A38, 28192A38, 29571A38, 31162A38, 32488A38, 35243AB39, 35295AB39, 36091AB39, 39464AB39, 40245AB39, 41317AB39, 26285A38, 27348A38, 29169A38, 30710A38, 31261A38, 32687A38, 35259AB39, 35359AB39, 36289AB39, 39031AB39, 41124AB39, 27236A38, 35051A38, 35839AB39, 40141AB39, 41702AB39
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Siemens Calibrator E; SMN 10309079 (2 pack), SMN 10321075 (6 pack), SMN 10335532 (6 pack Ref) fue retirado del mercado?

Sí. FDA publicó el retiro Z-0164-2015 el 12 de noviembre de 2014. Peligro reportado: Positive bias for Calibrator E lot kits ending in 38 and 39 with the ADVIA Centaur Systems Cortisol and Progesterone assays compared to the respective master curves.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter