Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0164-2026

IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida

Publicado el 5 de noviembre de 2025 · Inicio del retiro: 30 de septiembre de 2025

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
IMPRIS Imaging Inc
Distribuidor
Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
Lote
IMRIS Part Number(s): 120823-000, UDI/DI 00857534006806
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States