FDAFDA · Alimentos y medicamentos·N.º oficial Z-0193-2021
6.5" (17 cm) Appx 0.30 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave, NanoClave T-Connector, Purple Clamp, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709062525(17)250101(30)50(10)4558455; (01)10887709062525(17)241201(30)50(10)4537168; (01)10887709062525(17)241201(30)50(10)4493389. The high pressure extension set is a device used as conduit tubing to deliver fluids from an ~f electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices.
Publicado el 28 de octubre de 2020 · Inicio del retiro: 6 de julio de 2020
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- ICU Medical, Inc.
- Distribuidor
- Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
- Lote
- Lot Numbers:4558455, 4537168, 4493389.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.