Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, Catalog number 07250452001 / GTIN number 00365702700017 The Accu-Chek Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. The Accu-Chek Bolus Advisor, as a component of the Accu-Chek Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data
Publicado el 13 de diciembre de 2017 · Inicio del retiro: 8 de junio de 2017
Resumen
FDA registra el retiro Z-0197-2018 de dispositivos médicos de Clase II, publicado el 13 de diciembre de 2017 e iniciado el 8 de junio de 2017. Empresa responsable: Roche Diabetes Care, Inc.; Estado: Finalizado; Cantidad reportada: 1,134 bolus advisor features used; Lote: Software Versions 1.0.5, 1.0.15, 1.2.0, 1.2.2, 1.2.3, 2.0.0, 2.0.1, 2.1.0, and 2.1.1 for iOS and Android. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. After pairing a meter with the app for the first time, a customer may encounter the rare condition in which the countdown timer is not displayed and correction bolus advice is not available for the most recent, valid glucose reading. This same blood glucose value may become available for bolus advice calculation at a later time (countdown timer is displayed).
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Roche Diabetes Care, Inc.
- Distribuidor
- Worldwide Distribution - US (Nationwide) and Canada
- UPC
- 00365702700000
- Lote
- Software Versions 1.0.5, 1.0.15, 1.2.0, 1.2.2, 1.2.3, 2.0.0, 2.0.1, 2.1.0, and 2.1.1 for iOS and Android.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAccu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the3 de julio de 2024
- FDAAccu-Chek Aviva Plus Retail Strips 50 ct - intended for Blood Glucose Monitoring System Catalog Number:12 de febrero de 2020
- FDAAccu-Chek Aviva Plus Health Network Strips- intended for Blood Glucose Monitoring System Catalog Number:12 de febrero de 2020
- FDAAccu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is9 de octubre de 2019
- FDATrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM), Catalog Number 221261 Trypticase Soy Agar with 5% or 10%13 de diciembre de 2017
- FDABD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is designed for use with the BD Phoenix"13 de diciembre de 2017
Preguntas sobre este retiro
¿Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, Catalog number 07250452001 / GTIN number 00365702700017 The Accu-Chek Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. The Accu-Chek Bolus Advisor, as a component of the Accu-Chek Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data fue retirado del mercado?
Sí. FDA publicó el retiro Z-0197-2018 el 13 de diciembre de 2017. Peligro reportado: Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. After pairing a meter with the app for the first time, a customer may encounter the rare condition in which the countdown timer is not displayed and correction bolus advice is not available for the most recent, valid glucose reading. This same blood glucose value may become available for bolus advice calculation at a later time (countdown timer is displayed).
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter
¿Cómo sé si mi unidad está incluida en el retiro?
Compara los datos de tu producto con los del aviso. UPC: 00365702700000.