Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. Product REF Number: 08453071001
Publicado el 3 de julio de 2024 · Inicio del retiro: 13 de mayo de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Roche Diabetes Care, Inc.
- Distribuidor
- Nationwide and Worldwide
- Lote
- GTIN: 00365702729100 Serial Numbers: US: 92339920445, 92340117408, 92339920116, 92339955415, 92340120006, 92339094787, 92340116052,92340120057, 92339744998. Canada:93040305146
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States