Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0242-2021

11 IN (28cm) APPX 1.5ml EXT SET NanoClave, 6-Port NanoClave Manifold . 1 unit per pouch, 50 pouches per case. UDI:¿(01)10887709098937(17)241201(30)50(10)4493669;(01)10887709098937(17)250101(30)50(10)4558624. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Publicado el 28 de octubre de 2020 · Inicio del retiro: 6 de julio de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
ICU Medical, Inc.
Distribuidor
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Lote
Lot Numbers:4493669, 4558624.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States