FDAFDA · Alimentos y medicamentos·N.º oficial Z-0262-2020
POM Procedural Oxygen Masks, Endoscopic, Conscious Sedation, Bronchoscopy Mask, as follows: (1) POM Medical, LLC, REF 1001-MM, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA; (2) curaplex, REF 301-0318LT, Medium Concentration, Distributed by Sarnova HC, LLC's family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc., Dublin, OH; and (3) POM Medical, LLC REF 1001-MF, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA. All are packaged 1 mask/plastic pouch, 30 masks/case.
Publicado el 6 de noviembre de 2019 · Inicio del retiro: 11 de julio de 2018
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Complaints were received on the product, such as the re-breather bag would not fill/inflate or the oxygen line pops off after the oxygen is turned on due to a clogged port.
Qué hacer
Voluntary: Firm initiated · Telephone
Detalles del retiro
- Fabricante
- POM Medical LLC
- Distribuidor
- Distribution was made to CA, FL, MO, OH, and PA.
- Lote
- Model 1001-MM - Lot numbers 021418-021421, 032218-032221, 032618-032621, and 032818-032821; Model KE1001-MM, an internal number used by one of the distributors (same as 1001-MM) - Lot numbers 021418-021421 and 032818-032821; Model 301-0318LT - Lot numbers 022218-022221, 032218-032221, 032618-032621; and Model 1001-MF - Lot number 032818-032821.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.