epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Publicado el 13 de noviembre de 2024 · Inicio del retiro: 24 de septiembre de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Siemens Healthcare Diagnostics Inc
- Distribuidor
- US Nationwide. Canada.
- Lote
- UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States