Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0293-2026

Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Publicado el 29 de octubre de 2025 · Inicio del retiro: 12 de septiembre de 2025

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Distribuidor
Worldwide
Lote
all serial numbers in distribution
Clasificación de riesgo
Class II
Área (FDA)
device
País
Israel

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta