FDAFDA · Alimentos y medicamentos·N.º oficial Z-0403-2016
Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
Publicado el 16 de diciembre de 2015 · Inicio del retiro: 17 de noviembre de 2015
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Olympus Corporation of the Americas
- Distribuidor
- US Nationwide Distribution including these states: NE, IL, FL, TX, OH, PA, IN, CA, NY, AL, ND, GA, AK, MI, WA, MS, NJ, IA, AZ, and RI.
- Lote
- Serial numbers 622483, 622523, 622553, 622566, 622568, 622569, 622573, 622576, 622614, 622615, 622616, 622617, 622618, 622619, 622620, 622632, 622643, 622644, 622645, 622646, 622647, 622654, 622655, 622656, 622657, 622658, 622664, 622666, 622667, 622687, 622688, 622693, 622704, 622705, 622706, 622710, 622711, 622712, 622718, 622719, 622720, 622722, 622723, 622750, 622753, 622764, 622766, 622767, 622769, 622771, 622773, 622775, 622777, 622778, 622785, 622789, 622795, 622797, 622798, 622801, 622806, 623624, 623626, 623629, 623637, 623638, 623639, 623641, 623661, 623669, 623671, 623672, 623673, 623677, 623687, 623688, 623705, 623709, 623713, 623719, 623720, 623722, 623723, 623729, 623730, 623731, 623735, 623736, 623737, 623738, 623739, 623740, 623741, 623742, 623743, 623747, 623749, 623750, 623751, 623752, 623776, 623778, 623784, 623786, 623787, 623788, 623799, 623800, 623801, 623802, 624004, 624005, 624006, 624008, 624011, 624012, 624016, 624017, 624018, 624019, 624024, 624028, 624030, 624041, 624042, 624046, 624054, 624055, 624056, 624057, 624064, 624066, 624070, 624071, 624073, 624076, 624077, 624080, 624082, 624089, 624099, 624100, 624101, 624104, 624107, 624109, 624110, 624112, 624114, 624117, 624119, 624121, 624151, 624158, 624162, 624167, 624170, 624618, 624622, 624643, 624678, 624679, 624682, 624683, 624720, 622539, 622540, 622570, 622572, 623734, 623748, 624739, and 624741.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?