SIEMENS brand ARTISTE Linear Accelerator Linac systems with component: any photon unflat beam option (Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Martinez, CA
Publicado el 11 de diciembre de 2013 · Inicio del retiro: 23 de octubre de 2013
Resumen
FDA registra el retiro Z-0438-2014 de dispositivos médicos de Clase II, publicado el 11 de diciembre de 2013 e iniciado el 23 de octubre de 2013. Empresa responsable: Siemens Medical Solutions USA, Inc; Estado: Finalizado; Cantidad reportada: 26 units; Lote: Serial Numbers: 5364, 5419, 5640, 5737, 5823, 5830, 5166, 5342, 5416, 5444, 5479, 5485, 5628, 5745, 5746, 5762, 5763, 5764, 5772, 5781, 5783, 5797, 5815, 5819, 5820, 5822, 5829, 5837, 5839, 5888, 5896, 5911, 5937. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Siemens Medical Solutions USA, Inc
- Distribuidor
- Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.
- Lote
- Serial Numbers: 5364, 5419, 5640, 5737, 5823, 5830, 5166, 5342, 5416, 5444, 5479, 5485, 5628, 5745, 5746, 5762, 5763, 5764, 5772, 5781, 5783, 5797, 5815, 5819, 5820, 5822, 5829, 5837, 5839, 5888, 5896, 5911, 5937
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPRIMUS, ONCOR, ARTISTE Linac systems with component 550 TxT Treatment Table. TT-A, TT-D, TT-M or TT-S Table29 de enero de 2014
- FDASIMVIEW NT and SIMVIEW NT with IDI: October 01, 2012 Siemens ceased Simview distribution. Manufactured by:29 de enero de 2014
- FDASIEMENS brand ONCOR series Linear Accelerator Linac systems with component: any photon unflat beam11 de diciembre de 2013
- FDAC-Series: Clinac, Trilogy, Trilogy Tx. , Novalis high energy linear accelerators and UNIQUE single energy11 de diciembre de 2013
- FDAAssembly, Patient Side Manipulator on da Vinci Surgical Systems: da Vinci S Surgical System; da Vinci SI11 de diciembre de 2013
- FDAPRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT,4 de septiembre de 2013
Preguntas sobre este retiro
¿SIEMENS brand ARTISTE Linear Accelerator Linac systems with component: any photon unflat beam option (Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Martinez, CA fue retirado del mercado?
Sí. FDA publicó el retiro Z-0438-2014 el 11 de diciembre de 2013. Peligro reportado: A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter