PRIMUS, ONCOR, ARTISTE Linac systems with component 550 TxT Treatment Table. TT-A, TT-D, TT-M or TT-S Table tops. Manufactured by: SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse 19-21 Kemnath, GERMANY 95478 The intended use of the SIEMENS branded, ARTISTE", ONCOR" and PRIMUS" family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Publicado el 29 de enero de 2014 · Inicio del retiro: 5 de diciembre de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Update to correct the possibility for patients or operators to have their fingers trapped or pinched between the table top and the bottom frame of the 550 TxT Treatment Table. Combined with this update instruction is a second complaint issue related to an incident in which a cable inside of the TxT" Treatment Table has been damaged due to contacting a moving part. The resulting short-circuit cau
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Siemens Medical Solutions USA, Inc
- Distribuidor
- Worldwide Distribution: US (nationwide) and internationally to: Canada, Italy, P.R. China, Argentina, Pakistan, South Africa, Egypt, Iran, Vietnam, Indonesia, Brazil, Hungary, Bulgaria, India, Spain, Switzerland, Trinidad, Tobago, Italy, Algeria, Croatia, Turkey, Japan, Germany, Portugal, Poland, France, Sri Lanka, Russian Federation, France, Lebanon, Slovakia, Ukraine, Pakistan, and Korean Republic.
- Lote
- 550 TxT table - part number 07346534.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States