Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0507-2013

Siemens Healthcare BR Assay for CA 27.29 250 Test Kits 03896216 50 Test Kit 02419937 Ref Kit (US only) 10340081 Intended Use The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur and ADVIA Centaur XP systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage Ill breast cancer.

Publicado el 19 de diciembre de 2012 · Inicio del retiro: 14 de noviembre de 2012

Resumen

FDA registra el retiro Z-0507-2013 de dispositivos médicos de Clase II, publicado el 19 de diciembre de 2012 e iniciado el 14 de noviembre de 2012. Empresa responsable: Siemens Healthcare Diagnostics, Inc; Estado: Finalizado; Cantidad reportada: 250 test =1678 and 50 test =2660 and Ref. 1196; Lote: All lot codes. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Operating range of the ADVIA Centaur systems is 18®C to 30®C (64.4®F to 86.0®F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Siemens Healthcare Diagnostics, Inc
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
Lote
All lot codes
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Siemens Healthcare BR Assay for CA 27.29 250 Test Kits 03896216 50 Test Kit 02419937 Ref Kit (US only) 10340081 Intended Use The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur and ADVIA Centaur XP systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage Ill breast cancer. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0507-2013 el 19 de diciembre de 2012. Peligro reportado: Operating range of the ADVIA Centaur systems is 18®C to 30®C (64.4®F to 86.0®F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter