Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0522-2025

Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495

Publicado el 4 de diciembre de 2024 · Inicio del retiro: 1 de noviembre de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Kreatech Biotechnology Bv
Distribuidor
US States: FL, NY
Lote
UDI-DI: (01)08718858480241 Lot Number: 00088454
Clasificación de riesgo
Class II
Área (FDA)
device
País
Netherlands

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta