CoverStainer Slide Rack The CoverStainer Slide Rack is an accessory to the CoverStainer instrument. The purpose is to hold microscopic slides, during transfer through the staining process. The rack can hold up to 10 slides at a time, and is necessary for optimal CoverStainer instrument performance.
Publicado el 26 de diciembre de 2012 · Inicio del retiro: 26 de noviembre de 2012
Resumen
FDA registra el retiro Z-0563-2013 de dispositivos médicos de Clase II, publicado el 26 de diciembre de 2012 e iniciado el 26 de noviembre de 2012. Empresa responsable: Dako Denmark A/S; Estado: Finalizado; Cantidad reportada: 25 Units US (50-100 racks per unit) 106 total; Lote: The racks affected by this recall were distributed with the following CoverStainer instrument serial numbers in the U.S.: CS 100-10031 CS 100-10033 CS 100-10035 CS 100-10054 CS 100-10080 CS 100-10081 CS 100-10082 CS 100-10083 CS 100-10093 CS 100-10094 CS 100-10106 CS 100-10107 CS 100-10110 CS 100-10113 CS 100-10117 CS 100-10123 CS 100-10124 CS 100-10125 CS 100-10133 CS 100-10135 CS 100-10148 CS 100-10149 CS 100-10150 CS 100-10152 CS 100-10159. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Dako Denmark A/S
- Distribuidor
- Worldwide Distribution - USA (nationwide) and the following countries: Australia France Rep. of Korea Belgium Germany Russia Brazil Israel Saudi Arabia Canada Italy Spain Chile Japan Sweden China Mexico Taiwan Denmark Netherlands Turkey Ecuador Norway UK Finland Poland
- Lote
- The racks affected by this recall were distributed with the following CoverStainer instrument serial numbers in the U.S.: CS 100-10031 CS 100-10033 CS 100-10035 CS 100-10054 CS 100-10080 CS 100-10081 CS 100-10082 CS 100-10083 CS 100-10093 CS 100-10094 CS 100-10106 CS 100-10107 CS 100-10110 CS 100-10113 CS 100-10117 CS 100-10123 CS 100-10124 CS 100-10125 CS 100-10133 CS 100-10135 CS 100-10148 CS 100-10149 CS 100-10150 CS 100-10152 CS 100-10159
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Denmark
Retiros relacionados
- FDATitanium Tack Starter Kit, REF 400-270, Rx Only, non-sterile, BIOHORIZONS, BIRMINGHAM, AL 35244. Dental.26 de diciembre de 2012
- FDACase Carton: BD Oral Dispensing Syringe 1 mL Clear with Tip Cap BD, Franklin Lakes, NJ 07417 USA Made in USA26 de diciembre de 2012
- FDAAutoTac Delivery Handle, REF 400-200, Rx Only, non-sterile, BIOHORIZONS, BIRMINGHAM, AL 35244. Dental.26 de diciembre de 2012
- FDABleaseFocus Anesthesia Workstation and 700/900 Series Ventilators are anesthesia gas machine. In most cases,26 de diciembre de 2012
- FDAAdvantage Workstation Volume Share 2 (versionAW4.4_04) with Volume Viewer 3 and 3.1 (version7.0 to version7.526 de diciembre de 2012
- FDAThe Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer26 de diciembre de 2012
Preguntas sobre este retiro
¿CoverStainer Slide Rack The CoverStainer Slide Rack is an accessory to the CoverStainer instrument. The purpose is to hold microscopic slides, during transfer through the staining process. The rack can hold up to 10 slides at a time, and is necessary for optimal CoverStainer instrument performance. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0563-2013 el 26 de diciembre de 2012. Peligro reportado: The expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter