Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic® DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately controlled with medication. STN/GPi target system components: ¿ Kinetra® Model 7428 or Soletra® Model 7426 Neurostimulator ¿ Activa® PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead Medtronic DBS Therapy for Tremor - Unilateral thalamic stimulation of the ventral intermediate nucleus (VIM) using Medtronic DBS Therapy for Tremor is indicated for the suppression of tremor in the upper extremity. The system is intended for patients who are diagnosed with essential tremor or parkinsonian tremor not adequately controlled by medications and where the tremor constitutes a significant functional disability. VIM target system components: ¿ Soletra Model 7426 Neurostimulator ¿ Activa PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead.
Publicado el 26 de diciembre de 2012 · Inicio del retiro: 3 de diciembre de 2012
Resumen
FDA registra el retiro Z-0567-2013 de dispositivos médicos de Clase II, publicado el 26 de diciembre de 2012 e iniciado el 3 de diciembre de 2012. Empresa responsable: Medtronic Neuromodulation; Estado: Finalizado; Cantidad reportada: 140 units; Lote: NHU236350V to NHU236373V; NKN037257V to NKN037259V; NKN037372V to NKN037387V; NKN037476V to NKN037507V; NKN037692V to NKN037698V; NKN037772V to NKN037779V; NKN037790V to NKN037810V; NKN037812V to NKN037840V. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked. Approximately 14% of the 140 affected extension kits may contain a damaged radiopaque winged connector boot.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Medtronic Neuromodulation
- Distribuidor
- Worldwide Distribution -- USA (nationwide) including the states of CT, MI, NH, NY, OH, PA and VT., and the countries of AUSTRALIA, JAPAN, BELGIUM, FRANCE, GERMANY, ITALY, NETHERLANDS, POLAND, SPAIN, SWITZERLAND, TURKEY and UNITED KINGDOM.
- Lote
- NHU236350V to NHU236373V; NKN037257V to NKN037259V; NKN037372V to NKN037387V; NKN037476V to NKN037507V; NKN037692V to NKN037698V; NKN037772V to NKN037779V; NKN037790V to NKN037810V; NKN037812V to NKN037840V.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAInterStim SureScan MRI Leads 978A1 & 978B1: LEAD 978A128 ISTM MICRO 28CM SSMRI TINED 978A128 LEAD 978A133 ISTM28 de octubre de 2026
- FDAInterStim Percutaneous Extensions EXT 3560022 ISTM II PERC 3560022 EXT 3560030 ISTM MICRO PERC 356003028 de octubre de 2026
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted,3 de junio de 2026
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted,3 de junio de 2026
- FDAThe Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer26 de diciembre de 2012
- FDAMedline Angio Drape Pack Radiology-LF; a sterile custom single patient prescription procedure pack, including26 de diciembre de 2012
Preguntas sobre este retiro
¿Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic® DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately controlled with medication. STN/GPi target system components: ¿ Kinetra® Model 7428 or Soletra® Model 7426 Neurostimulator ¿ Activa® PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead Medtronic DBS Therapy for Tremor - Unilateral thalamic stimulation of the ventral intermediate nucleus (VIM) using Medtronic DBS Therapy for Tremor is indicated for the suppression of tremor in the upper extremity. The system is intended for patients who are diagnosed with essential tremor or parkinsonian tremor not adequately controlled by medications and where the tremor constitutes a significant functional disability. VIM target system components: ¿ Soletra Model 7426 Neurostimulator ¿ Activa PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0567-2013 el 26 de diciembre de 2012. Peligro reportado: Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked. Approximately 14% of the 140 affected extension kits may contain a damaged radiopaque winged connector boot.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter