SOL SYS 8IN CALC 1.5/19.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
Publicado el 26 de diciembre de 2012 · Inicio del retiro: 3 de diciembre de 2012
Resumen
FDA registra el retiro Z-0578-2013 de dispositivos médicos de Clase II, publicado el 26 de diciembre de 2012 e iniciado el 3 de diciembre de 2012. Empresa responsable: DePuy Orthopaedics, Inc.; Estado: Finalizado; Cantidad reportada: 1732 total all product codes; Lote: Product Code: 157215195 and Lot codes: 126367, A34FG1000, A63GS1000, B19HM1000, B3DE61000, B46AJ1000, BY6G81000, C4YBH1000, C5FH81000, CB9L21000, CH7EH1000, CP8NJ1000, D83FY1000, DG1PB1000, X24EW1000, X24EX1000, X24EY1000, X24EYB000, X3DC51000, X5SB31000, X5SB3A000, YA6GT1000, YF6AE1000, YF6AEA000, Z5VAN1000, and Z5VANB000. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- DePuy Orthopaedics, Inc.
- Distribuidor
- Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR
- Lote
- Product Code: 157215195 and Lot codes: 126367, A34FG1000, A63GS1000, B19HM1000, B3DE61000, B46AJ1000, BY6G81000, C4YBH1000, C5FH81000, CB9L21000, CH7EH1000, CP8NJ1000, D83FY1000, DG1PB1000, X24EW1000, X24EX1000, X24EY1000, X24EYB000, X3DC51000, X5SB31000, X5SB3A000, YA6GT1000, YF6AE1000, YF6AEA000, Z5VAN1000, and Z5VANB000.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total Knee10 de abril de 2024
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 15176021229 de marzo de 2023
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 15176031229 de marzo de 2023
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation27 de julio de 2022
- FDASOL SYS 8IN CALC 1.5/18.0 Packaging: Product is packed within a poly protector, with protection added as26 de diciembre de 2012
- FDASOL SYS F 9 CALC L 13.5MM LG Packaging: Product is packed within a poly protector, with protection added as26 de diciembre de 2012
Preguntas sobre este retiro
¿SOL SYS 8IN CALC 1.5/19.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0578-2013 el 26 de diciembre de 2012. Peligro reportado: An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter