Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system

Publicado el 11 de diciembre de 2024 · Inicio del retiro: 8 de agosto de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

Qué hacer

FDA Mandated · Letter

Detalles del retiro

Fabricante
Orthoscan, Inc.
Distribuidor
U.S. (including Puerto Rico).
Lote
UDI: EZIEZIEHMVISIONRFD3D1. Serial Numbers: 21118, 21212, 22040, 22710, 22764, 23247, 23248, 23255, 23257, 23258, 23259, 23260, 23327, 23328, 23329, 23456, 23457, 23458, 23459, 23653, 23654, 23655, 23656, 23657, 23666, 23667, 23668, 23669, 23670, 23671, 23672, 23673, 23674, 23723, 23740, 23744, 23745, 23746, 23747, 23748, 23749, 23750, 23751, 23752, 23753, 23775, 23801, 23802, 23828, 23849.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States