ADULT/CHILD Radiotranslucent Electrode, Part number T100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.
Publicado el 31 de diciembre de 2014 · Inicio del retiro: 11 de noviembre de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- Heart Sync, Inc.
- Distribuidor
- Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.
- Lote
- All Lots. Catalog / Part Number T100AC-PHILIPS
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States