FDAFDA · Alimentos y medicamentos·N.º oficial Z-0832-2015
HemosIL PT-Fibrinogen HS PLUS, IVD --- Instrumentation Laboratory IL Company. A high sensitivity thromboplastin reagent based on recombinant human tissue factor (RTF) for the quantitative determination in human citrated plasma of Prothrombin Time (PT) and Fibrinogen on IL Coagulation Systems. The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy (OAT).
Publicado el 31 de diciembre de 2014 · Inicio del retiro: 21 de noviembre de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Some vials of HemosIL PT-Fibrinogen HS PLUS, Part No. 0008469810 (various lots manufactured prior to June 2014) exhibit a yellow/brownish color after reconstitution (yellow color or darker) instead of the typical white to off-white color. Vials exhibiting this yellow/brownish color were observed to cause prolonged prothrombin (PT) clotting times.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Instrumentation Laboratory Co.
- Distribuidor
- Worldwide Distribution: US (nationwide) including Puerto Rico; and countries of: ALBANIA, ALGERIA, ANDORRA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BELARUS, BRAZIL, BRUNEI, BULGARIA, CHILE, CHINA, COLUMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FRANCE, GABON, GEORGIA, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, TURKEY, UAE, UGANDA, UKRAINE, UNITED KINGDOM, AND URUGUAY.
- Lote
- Lots: N1122175, exp. date 11/30/2014; N0132936, exp. date 1/31/2015; N0233302, exp. date 2/28/2015; N0333382, exp. date 3/31/2015; E0333382, exp. date 3/31/2015; N0333626, exp. date 3/31/2015; N0433932, exp. date 4/30/2015; N0634953, exp. date 6/30/2015; N0735317, exp. date 5/31/2015; N0735599, exp. date 7/31/ 2015; N0836106, exp. date 8/31/2015; N0936197, exp. date 9/30/2015; N0349325, exp. date 12/31/2015; N1222324, exp. date 12/31/2015; N0148094, exp. date 12/31/2015; N0248652, exp. date 2/29/2016; N0249017, exp. date 2/29/2016; N0449790, exp. date 4/30/2016; N0440009, exp. date 4/30/2016.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?