Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0598-2015

PKG, FIXATION CLAMP 5MM WITH RATCHET, P/N 0250080130 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Stryker Endoscopy
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lote
542650 744316 746309 746656 840349 845864 940057 940150 940431 942656 944708 945457 1040794 1041288 1043107 1043115 1045071 1046601 1140168 1142891 1143157 1143970 1144494 1144654 1146465 1146836 1147098 1241415 1242117 1242944 1245442 1245450 1340115 1341560 1341562 1343425 1344136 1441095 790382
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States