Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0598-2016

Brainlab Cranial Navigation System: An Image Guided Surgery System / Stereotactic. Radiology Departments. The BrainLAB Cranial IGS System is intended to be an intra-operative image guided localization system to enable minimally invasive surgery.

Publicado el 27 de enero de 2016 · Inicio del retiro: 22 de abril de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Software Error: The effect of setup on overall navigation accuracy could potentially intensify small inaccuracies arising from individual steps of a complex navigation procedure that may cause an inaccurate display of instruments by the navigation system compared to the actual patient anatomy.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Brainlab AG
Distribuidor
Worldwide Distribution: US (Nationwide) including Puerto Rico; and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Latvia, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, and Vietnam.
Lote
1) Model 22035 , VVC NAVIGATION SOFTWARE CRANIAL 5.0 , Serial Numbers: 22035-103911-004 , 22035-103953-001 , 22035-103977-001 , 22035-103982-002 , 22035-103987-001 , 22035-103988-001 , 22035-104566-001 ; 2) Model 22040 , VV NAVIGATION SOFTWARE CRANIAL 4.0 , Serial Numbers: 22040/18820 , 22040/21370 , 22040/26909 , 22040/36997 , 22040-103675-006 , 22040-103687-009 , 22040-103710-001 , 22040-103942-008 , 22040-103974-011 , 22040-104067-001 , 22040-104072-001 , 22040-104114-001 , 22040-104148-001 , 22040-104151-001 , 22040-104505-001 , 22040-104542-001 , 22040-104545-001 , 22040-104614-005 , 22040-123772-001 ; 3) Model 22040-02 , VV NAV SW CRANIAL 7.8.1 (ANY PLATFORM) , Serial Number: 000000002600509001 ; 4) Model 22040-11 , UPDATE VV NAV SW CRANIAL 7.9.1 FROM 7.9 , Serial Numbers: 000000007665811001 , 000000007801111004 , 000000007801111005 , 000000007801111006 , 000000007801111007 , 000000007801111008 , 000000007801111009 , 000000007808711004 , 000000007808711007 , 000000007808711008 , 000000007837611010 , 000000007847411007 , 000000007865211004 , 000000007865211005 , 000000007942211004 , 000000007942211008 , 000000007942211010 , 000000007950411001 , 000000007950411004 , 000000007950411007 , 000000007950411010 , 000000007956611005 , 000000007956611010 , 000000007956611017 , 000000007965011001 , 000000007965011006 , 000000007965011007 , 000000007980111001 , 000000007980111002 , 000000007980111003 , 000000007980111004 , 000000007980111005 , 000000007980111006 , 000000007980111007 , 000000008000111003 , 000000008000111006 , 000000008000111007 , 000000008012411001 , 000000008012411003 , 000000008012411008 , 000000008026011002 , 000000008026011003 , 000000008036811002 , 000000008036811006 , 000000008036811007 , 000000008036811008 , 000000008036911002 , 000000008036911004 , 000000008036911005 , 000000008046111002 , 000000008046111005 , 000000008073511003 , 000000008073511007 , 000000008086811001 , 000000008106711001 , 000000008114111004 , 000000008129711004 , 000000008129711005 , 000000008129711006 , 000000008133311001 , 000000008133311002 , 000000008143111002 , 000000008143111003 , 000000008143111004 , 000000008157411005 , 000000008157411007 , 000000008167811005 , 000000008167811008 , 000000008167811012 , 000000008167811014 , 000000008167811015 , 000000008167811016 , 000000008180811001 , 000000008180811008 , 000000008180811009 , 000000008180811010 , 000000008180811011 , 000000008181011002 , 000000008208411002 , 000000008230111001 , 000000008230111002 , 000000008230111009 , 000000008230111013 , 000000008230111015 , 000000008230111022 , 000000008230111024 , 000000008230111025 , 000000008230211001 , 000000008230211002 , 000000008230211008 , 000000008230211017 , 000000008230211020 , 000000008230211025 , 000000008230211030 , 000000008230311006 , 000000008230311014 , 000000008230311015 , 000000008230311018 , 000000008230411003 , 000000008230411005 , 000000008230411009 , 000000008230411010 , 000000008230411012 , 000000008230411016 , 000000008230411021 , 000000008230411024 , 000000008230411029 , 000000008230411030 , 000000008230411031 , 000000008230411032 , 000000008242111002 , 000000008242111003 , 000000008242111005 , 000000008242611001 , 000000008242611002 , 000000009168012001 ; 5) Model 22040B , VV NAVIGATION SOFTWARE CRANIAL7.5.0 , Serial Numbers: 05-48916-22040B , 05-51803-22040B , 05-52149-22040B , 22040B/46911 ; 6) Model 22040C , VV NAVIGATION SOFTWARE CRANIAL 7.6.0 , Serial Numbers: 05-63659-22040C , 05-63669-22040C , 05-66754-22040C , 06-79914-22040C ; 7) Model 22040D , VV NAVIGATION SW CRANIAL 7.7 , Serial Numbers: 06-84302-22040D , 06-85635-22040D , 06-88801-22040D ; 8) Model 22040E , VV NAVIGATION SW CRANIAL 7.8 , Serial Numbers: 06-101594-22040E , 06-103819-22040E , 06-108132-22040E , 07-102189-22040E , 07-106354-22040E , 07-108464-22040E , 07-109071-
Clasificación de riesgo
Class I
Área (FDA)
device
País
Germany

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta