The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network. This software system will not be the software directly used to run, steer, or control any specific laboratory analyzer or equipment. It is intended for use in laboratory facilities, including a central laboratory or in a multiple laboratory environment servicing satellite laboratories, reference laboratories, clinics, etc. Centricity Laboratory System is not intended for direct patient contact. The Centricity Laboratory System product is designed to facilitate the general clinical, anatomic pathology and cytology laboratory workflow such as order entry, results entry, instrument interfacing, results reporting and patient record retrieval. It integrates with other Hospital Information Systems (HIS) through HL7 interface by order entering and also report processing. The Centricity Laboratory System is intended to interface with various lab results. Centricity Laboratory System is intended to be used by laboratory technologists, technicians, analyst and other trained/site authorized system users
Publicado el 8 de enero de 2014 · Inicio del retiro: 12 de marzo de 2013
Resumen
FDA registra el retiro Z-0620-2014 de dispositivos médicos de Clase II, publicado el 8 de enero de 2014 e iniciado el 12 de marzo de 2013. Empresa responsable: GE Healthcare It; Estado: Finalizado; Cantidad reportada: 5 units; Lote: Software Versions: 3.3, 4.0, 4.1. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
GE Healthcare is aware of a potential safety issue with the use of GE Centricity Laboratory Instrument Interface (IF) where free-text sent from IM is not being transcribed into free-text in iNET. The instrument may upload results to IF/iNET when order/request is not yet in IF/iNET. In such a situation, test level supporting free-text sent from the instrument is discarded (not captured).
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Healthcare It
- Distribuidor
- Worldwide Distribution - US including MN and Internationally to India, Australia, Canada, and England.
- Lote
- Software Versions: 3.3, 4.0, 4.1
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an information7 de septiembre de 2016
- FDAThe Centricity Laboratory System is intended to be an information system designed to support the clinical and5 de febrero de 2014
- FDAMerge Mammo is a multi-modality, vendor-neutral digital mammography workstation that enables imaging centers22 de enero de 2014
- FDACarisoprodol Tablets, USP, CIV, 350 mg, 500 count bottle, Rx only. Manufactured by Watson Pharma Private Ltd,15 de enero de 2014
- FDAAdvanDx Gram-Negative QuickFISH BC. In-Vitro Diagnostic for the identification of Enterobacteriacae Catalog8 de enero de 2014
- FDACentricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare27 de noviembre de 2013
Preguntas sobre este retiro
¿The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network. This software system will not be the software directly used to run, steer, or control any specific laboratory analyzer or equipment. It is intended for use in laboratory facilities, including a central laboratory or in a multiple laboratory environment servicing satellite laboratories, reference laboratories, clinics, etc. Centricity Laboratory System is not intended for direct patient contact. The Centricity Laboratory System product is designed to facilitate the general clinical, anatomic pathology and cytology laboratory workflow such as order entry, results entry, instrument interfacing, results reporting and patient record retrieval. It integrates with other Hospital Information Systems (HIS) through HL7 interface by order entering and also report processing. The Centricity Laboratory System is intended to interface with various lab results. Centricity Laboratory System is intended to be used by laboratory technologists, technicians, analyst and other trained/site authorized system users fue retirado del mercado?
Sí. FDA publicó el retiro Z-0620-2014 el 8 de enero de 2014. Peligro reportado: GE Healthcare is aware of a potential safety issue with the use of GE Centricity Laboratory Instrument Interface (IF) where free-text sent from IM is not being transcribed into free-text in iNET. The instrument may upload results to IF/iNET when order/request is not yet in IF/iNET. In such a situation, test level supporting free-text sent from the instrument is discarded (not captured).
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter